APPLICATION NOTE

Metals in Sewage Sludge

SUMMARY

The application note summarizes the digestion of SQC001S, a sewage sludge Certified Reference Material using ColdBlock Digestion Pro Series Technology.

Instrument:ColdBlock CBM sample digester, chiller, ICP-OES, ICP-MS (with CRC technology)
Published:December 2024
Digestion Time:45 minutes
Acid Used:HNO3 , HCl & H2O2
Average ColdBlock Recovery vs. CRM:
  • 103% Arsenic

  • 96% Mercury

  • 98% Lead

METHODOLOGY

  1. Chiller temperature was set to -5°C
  2. 0.5g of each sample was weighed and placed into a ColdBlock™ Digestion vessel
  3. 10mL of 1:1 HNO3 + 5mL 1:1 HCl was added, and the samples were digested at 70% power for 5 minutes
  4. 10mL HNO3 was added and digested again at 60% power for 20 minutes (repeat this step if fuming continues)
  5. 3mL of 30% H2O2 was added slowly, and digested again at 60% for 10 minutes
  6. Add 5mL HCl and digest a final time at 60% power for 10 minutes
  7. Samples were cooled and bulked to 50mL using ultrapure water

DISCUSSION

  • This method yielded data comparable to EPA 3050b standards (with improved recovery of antimony) in just 45
  • The addition of HCl in the final step of the digestion helps keep certain elements such as antimony and silver solubilized
  • This method is not appropriate for standard ICP-MS instruments that are not equipped with Collision/Reaction Cell technology, for more detailed information please contact ColdBlock at info@coldblock.ca

Metals in Sewage Sludge, CRM, SQC001S, after bulk up to 50mL

Results

Metals in Sewage Sludge

Method:

0.5g

10mL 1:1 HNO3 + 5mL 1:1 HCl, digest at 70% for 5 minutes, then add 10mL HNO3 and digest at 60% for 20 minutes (repeat this step if fuming continues) Add 3mL of 30% H2O2, slowly and digest again

@ 60% for 10 minutes, then add 5mL HCl and digest again @ 60% for 10 minutes

5.Cool and bulk to 50mL with 2% HNO3

Element

Certified Value

+/-

Sample A

Sample B

Sample C

Sample D

Sample E

Average (ppm)

Stdev

% RSD

% Recovery vs certified value

Ag (ppm)

42.3

5.5

37.2

38.4

37.4

41.2

44.2

39.7

2.99

7.5%

94%

Al (ppm)

18800

3120

17907

19519

18208

17571

18164

18274

741

4.1%

97%

As (ppm)

282

35.9

284

294

293

271

316

292

16.5

5.6%

103%

Ba (ppm)

430

40.7

425

449

435

389

412

422

22.9

5.4%

98%

Be (ppm)

45.1

4.8

48.9

48.5

48.8

48.8

48.1

48.6

0.327

0.7%

108%

Ca (ppm)

66200

8280

68389

71253

65949

67212

67194

67999

2013

3.0%

103%

Cd (ppm)

64.4

5.2

59

60.5

62.9

62.9

65.6

62.2

2.53

4.1%

97%

Co (ppm)

49.3

4.39

48.4

47.4

45.9

48.6

53

48.7

2.65

5.4%

99%

Cr (ppm)

116

14.6

121

119

121

123

137

124

7.29

5.9%

107%

Cu (ppm)

416

33.8

416

391

396

423

485

422

37.6

8.9%

101%

Fe (ppm)

33900

5630

31232

32641

32119

31724

36715

32886

2202

6.7%

97%

Hg (ppm)

6.06

1.44

5.91

5.62

5.91

5.77

5.96

5.83

0.139

2.4%

96%

K (ppm)

4330

161

3991

3968

4447

4062

4484

4190

254

6.1%

97%

Mg (ppm)

13200

1110

14243

14001

13618

13184

14129

13835

433

3.1%

105%

Mn (ppm)

1310

350

1130

1550

1191

1340

1246

1291

164

12.7%

99%

Mo (ppm)

102

17

95.2

96.6

103.6

107.5

107.6

102

5.91

5.8%

100%

Na (ppm)

3100

304

2895

2967

2783

2995

3066

2941

107.6

3.7%

95%

Ni (ppm)

106

6.42

103.2

108.5

106.4

104

109.7

106

2.80

2.6%

100%

Pb (ppm)

126

10.5

121

121

125

119

130

123

4.38

3.6%

98%

Sb (ppm)

70.8

24

66.3

66.8

69.1

69.0

75.9

69.4

3.84

5.5%

98%

Se (ppm)

146

18.6

137

144

151

154

145

146

6.61

4.5%

100%

Tl (ppm)

40.9

10.9

38.2

39.1

38

37.7

44.1

39.4

2.67

6.8%

96%

V (ppm)

105

15.8

104

104

96.8

92.8

107

101

5.89

5.8%

96%

Zn (ppm)

913

82.8

905

885

863

881

1002

907

55.1

6.1%

99%

Click here to download the Application Note in PDF format.

Join our monthly e-newsletter for expert tips, updates, and exclusive content delivered straight to your inbox.

ColdBlock Monthly Newsletter

Simply enter your email address below to stay on top of the latest innovations and expertise from ColdBlock.

ColdBlock Technologies Inc.
4551 Zimmerman Ave
Niagara Falls, ON L2E 3M5
info@coldblock.ca.