SUMMARY
The application note summarizes the digestion of SQC001S, a sewage sludge Certified Reference Material using ColdBlock Digestion Pro Series Technology.
Instrument: | ColdBlock CBM sample digester, chiller, ICP-OES, ICP-MS (with CRC technology) |
Published: | December 2024 |
Digestion Time: | 45 minutes |
Acid Used: | HNO3 , HCl & H2O2 |
Average ColdBlock Recovery vs. CRM: |
|
Metals in Sewage Sludge, CRM, SQC001S, after bulk up to 50mL
Metals in Sewage Sludge |
|||||||||||
Method: |
0.5g |
10mL 1:1 HNO3 + 5mL 1:1 HCl, digest at 70% for 5 minutes, then add 10mL HNO3 and digest at 60% for 20 minutes (repeat this step if fuming continues) Add 3mL of 30% H2O2, slowly and digest again @ 60% for 10 minutes, then add 5mL HCl and digest again @ 60% for 10 minutes 5.Cool and bulk to 50mL with 2% HNO3 |
|||||||||
Element |
Certified Value |
+/- |
Sample A |
Sample B |
Sample C |
Sample D |
Sample E |
Average (ppm) |
Stdev |
% RSD |
% Recovery vs certified value |
Ag (ppm) |
42.3 |
5.5 |
37.2 |
38.4 |
37.4 |
41.2 |
44.2 |
39.7 |
2.99 |
7.5% |
94% |
Al (ppm) |
18800 |
3120 |
17907 |
19519 |
18208 |
17571 |
18164 |
18274 |
741 |
4.1% |
97% |
As (ppm) |
282 |
35.9 |
284 |
294 |
293 |
271 |
316 |
292 |
16.5 |
5.6% |
103% |
Ba (ppm) |
430 |
40.7 |
425 |
449 |
435 |
389 |
412 |
422 |
22.9 |
5.4% |
98% |
Be (ppm) |
45.1 |
4.8 |
48.9 |
48.5 |
48.8 |
48.8 |
48.1 |
48.6 |
0.327 |
0.7% |
108% |
Ca (ppm) |
66200 |
8280 |
68389 |
71253 |
65949 |
67212 |
67194 |
67999 |
2013 |
3.0% |
103% |
Cd (ppm) |
64.4 |
5.2 |
59 |
60.5 |
62.9 |
62.9 |
65.6 |
62.2 |
2.53 |
4.1% |
97% |
Co (ppm) |
49.3 |
4.39 |
48.4 |
47.4 |
45.9 |
48.6 |
53 |
48.7 |
2.65 |
5.4% |
99% |
Cr (ppm) |
116 |
14.6 |
121 |
119 |
121 |
123 |
137 |
124 |
7.29 |
5.9% |
107% |
Cu (ppm) |
416 |
33.8 |
416 |
391 |
396 |
423 |
485 |
422 |
37.6 |
8.9% |
101% |
Fe (ppm) |
33900 |
5630 |
31232 |
32641 |
32119 |
31724 |
36715 |
32886 |
2202 |
6.7% |
97% |
Hg (ppm) |
6.06 |
1.44 |
5.91 |
5.62 |
5.91 |
5.77 |
5.96 |
5.83 |
0.139 |
2.4% |
96% |
K (ppm) |
4330 |
161 |
3991 |
3968 |
4447 |
4062 |
4484 |
4190 |
254 |
6.1% |
97% |
Mg (ppm) |
13200 |
1110 |
14243 |
14001 |
13618 |
13184 |
14129 |
13835 |
433 |
3.1% |
105% |
Mn (ppm) |
1310 |
350 |
1130 |
1550 |
1191 |
1340 |
1246 |
1291 |
164 |
12.7% |
99% |
Mo (ppm) |
102 |
17 |
95.2 |
96.6 |
103.6 |
107.5 |
107.6 |
102 |
5.91 |
5.8% |
100% |
Na (ppm) |
3100 |
304 |
2895 |
2967 |
2783 |
2995 |
3066 |
2941 |
107.6 |
3.7% |
95% |
Ni (ppm) |
106 |
6.42 |
103.2 |
108.5 |
106.4 |
104 |
109.7 |
106 |
2.80 |
2.6% |
100% |
Pb (ppm) |
126 |
10.5 |
121 |
121 |
125 |
119 |
130 |
123 |
4.38 |
3.6% |
98% |
Sb (ppm) |
70.8 |
24 |
66.3 |
66.8 |
69.1 |
69.0 |
75.9 |
69.4 |
3.84 |
5.5% |
98% |
Se (ppm) |
146 |
18.6 |
137 |
144 |
151 |
154 |
145 |
146 |
6.61 |
4.5% |
100% |
Tl (ppm) |
40.9 |
10.9 |
38.2 |
39.1 |
38 |
37.7 |
44.1 |
39.4 |
2.67 |
6.8% |
96% |
V (ppm) |
105 |
15.8 |
104 |
104 |
96.8 |
92.8 |
107 |
101 |
5.89 |
5.8% |
96% |
Zn (ppm) |
913 |
82.8 |
905 |
885 |
863 |
881 |
1002 |
907 |
55.1 |
6.1% |
99% |
Click here to download the Application Note in PDF format.