SUMMARY
This application note is for the digestion of OREAS-353, a Lead Concentrate.
| Instrument: | ColdBlock CB15S sample digester technology, chiller, ICP-OES, ICP-MS  | 
| Published: | August 2022 | 
| Digestion Time: | 20 Minutes | 
| Acid Used: | Reverse Aqua Regia | 
| Average ColdBlock Recovery vs. CRM: | 
  | 
OREAS 353 is a certified reference material (CRM) prepared from a Pb-Ag concentrate sample sourced from South32 Ltd’s Cannington mine plant. The stratabound, metasediment hosted (Broken Hill Type) deposit is located ~200 kms south-east of Mount Isa in north- west Queensland, Australia.
| 
     Table 1: Results of Reverse Aqua Regia digestion  | 
    ||||||
| 
     Elements  | 
    
     Expected Value  | 
    
     ColdBlock Value 1  | 
    
     ColdBlock Value 2  | 
    
     Average  | 
    
     % RSD  | 
    
     % Recovery  | 
    
| 
     Ag (ppm)  | 
    
     2184  | 
    
     1987  | 
    
     1969  | 
    
     1978  | 
    
     0.5%  | 
    
     91%  | 
    
| 
     Al (ppm)  | 
    
     2040  | 
    
     794  | 
    
     640  | 
    
     717  | 
    
     10.7%  | 
    
     35%  | 
    
| 
     Ca (ppm)  | 
    
     2360  | 
    
     2411  | 
    
     2291  | 
    
     2351  | 
    
     2.6%  | 
    
     100%  | 
    
| 
     Cd (ppm)  | 
    
     205  | 
    
     198  | 
    
     206  | 
    
     202  | 
    
     2.0%  | 
    
     99%  | 
    
| 
     Co (ppm)  | 
    
     48.8  | 
    
     36.8  | 
    
     50.0  | 
    
     43.4  | 
    
     15.2%  | 
    
     89%  | 
    
| 
     Cu (ppm)  | 
    
     3120  | 
    
     2774  | 
    
     2790  | 
    
     2782  | 
    
     0.3%  | 
    
     89%  | 
    
| 
     Fe (wt.%)  | 
    
     5.35  | 
    
     4.81  | 
    
     4.81  | 
    
     4.81  | 
    
     0.0%  | 
    
     90%  | 
    
| 
     Mg (wt.%)  | 
    
     0.685  | 
    
     0.205  | 
    
     0.206  | 
    
     0.206  | 
    
     0.2%  | 
    
     30%  | 
    
| 
     Mn (wt.%)  | 
    
     0.184  | 
    
     0.147  | 
    
     0.147  | 
    
     0.147  | 
    
     0.0%  | 
    
     80%  | 
    
| 
     Pb (wt.%)  | 
    
     61.26  | 
    
     57.89  | 
    
     60.6  | 
    
     59.25  | 
    
     2.3%  | 
    
     97%  | 
    
| 
     S (wt.%)  | 
    
     15.13  | 
    
     13.66  | 
    
     14.33  | 
    
     14.00  | 
    
     2.4%  | 
    
     92%  | 
    
| 
     Sb (ppm)  | 
    
     1746  | 
    
     606  | 
    
     932  | 
    
     769  | 
    
     21.2%  | 
    
     44%  | 
    
| 
     Zn (wt.%)  | 
    
     4.14  | 
    
     4.12  | 
    
     4.07  | 
    
     4.10  | 
    
     0.6%  | 
    
     99%  | 
    
| 
 Table 2 – Results of adding HF, showing improved recoveries of Al, Co, Fe, Mg, Mn, Sb  | 
||||||
| 
 Elements  | 
 Expected Value  | 
 ColdBlock Value 1  | 
 ColdBlock Value 2  | 
 Average  | 
 % RSD  | 
 % Recovery  | 
| 
 AI (ppm)  | 
 2040  | 
 2049  | 
 1677  | 
 1863  | 
 0.1%  | 
 91%  | 
| 
 Co (ppm)  | 
 48.8  | 
 49.1  | 
 51.6  | 
 50.4  | 
 2.5%  | 
 103%  | 
| 
 Fe (wt.%)  | 
 5.35  | 
 5.26  | 
 4.87  | 
 5.07  | 
 0.3%  | 
 95%  | 
| 
 Mg (wt.%)  | 
 0.685  | 
 0.681  | 
 0.650  | 
 0.666  | 
 2.3%  | 
 97%  | 
| 
 Mn (wt.%)  | 
 0.184  | 
 0.181  | 
 0.184  | 
 0.183  | 
 0.8%  | 
 99%  | 
| 
 Sb (ppm)  | 
 1746  | 
 1738  | 
 1676  | 
 1707  | 
 1.8%  | 
 98%  | 
Click here to download the Application Note in PDF format.