APPLICATION NOTE

Contaminated Soil, SS-1

SUMMARY

The application note summarizes the digestion of SS-1 a contaminated soil Certified Reference Material using ColdBlock Digestion Pro Series Technology.

Instrument:ColdBlock CBM sample digester, chiller, ICP-OES, ICP-MS (with CRC technology)
Published:December 2024
Digestion Time:45 minutes
Acid Used:HNO3, HCl & H2O2
Average ColdBlock Recovery vs. CRM:
  • 99% Antimony
  • 96% Silver
  • 101% Lead

METHODOLOGY

  1. Chiller temperature was set to -5°C
  2. 0.5g of each sample was weighed and placed into a ColdBlock™ Digestion vessel
  3. 10mL of 1:1 HNO3 + 5mL 1:1 HCl was added, and the samples were digested at 70% power for 5 minutes
  4. 10mL HNO3 was added and digested again at 60% power for 20 minutes (repeat this step if fuming continues)
  5. 3mL of 30% H2O2 was added slowly, and digested again at 60% for 10 minutes
  6. Add 5mL HCl and digest a final time at 60% power for 10 minutes
  7. Samples were cooled and bulked to 50mL using ultrapure water

DISCUSSION

  • This method yielded data comparable to EPA 3050b standards (with improved recovery of antimony) in just 45
  • The addition of HCl in the final step of the digestion helps keep certain elements such as antimony and silver solubilized
  • This method is not appropriate for standard ICP-MS instruments that are not equipped with Collision/Reaction Cell technology, for more detailed information please contact ColdBlock at info@coldblock.ca

Contaminated Soil, SS-1 after bulk up to 50mL

Results

Contaminated Soil, SS-1

Method:

0.5g

10mL 1:1 HNO3 + 5mL 1:1 HCl, digest at 70% for 5 minutes, then add 10mL HNO3 and digest at 60% for 20 minutes (repeat this step if fuming continues) Add 3mL of 30% H2O2, slowly and digest again @ 60% for 10 minutes, then add 5mL HCl and digest again @ 60% for 10 minutes.Cool and bulk to 50mL with 2% HNO3

Element

Certified Value

+/-

Sample A

Sample B

Sample C

Sample D

Sample E

Average (ppm)

Stdev

% RSD

% Recovery vs certified value

Ag

0.88

0.03

0.84

0.84

0.89

0.84

0.82

0.85

0.03

3.1%

96%

Al

12163

410

12077

12634

11519

11542

11985

11951

458

3.8%

98%

As

20.7

1

19.3

20.6

19.6

19.9

20.3

19.9

0.522

2.6%

96%

Ba

464

16

445

473

456

431

436

448

16.8

3.8%

97%

Be

0.48

0.05

0.49

0.46

0.47

0.48

0.49

0.48

0.013

2.7%

100%

Ca

50265

1213

49491

51721

47580

47290

49195

49055

1775

3.6%

98%

Cd

3.2

0.2

3.5

3.4

3.2

3.4

3.0

3.3

0.20

6.1%

103%

Co

12.9

0.4

12.8

13

12.4

12.1

12.9

12.6

0.38

3.0%

98%

Cr

103

5

98.1

103

101

97.5

98.8

100

2.28

2.3%

97%

Cu

403

10

399

427

384

370

385

393

21.6

5.5%

98%

Fe

72000

2273

69216

71857

67438

68507

68635

69131

1654

2.4%

96%

Hg

0.41

0.02

0.38

0.40

0.39

0.41

0.38

0.39

0.01

3.3%

96%

K

2232

150

2493

2576

2438

2523

2468

2500

53

2.1%

112%

Mg

9690

230

9275

9325

9206

9274

9547

9325

131

1.4%

96%

Mn

737

19

708

738

691

687

719

709

21

3.0%

96%

Mo

6.8

0.3

7.3

7.4

6.5

7.5

6.9

7.1

0.41

5.8%

105%

Na

650

64

637

641

582

593

615

614

26.1

4.3%

94%

Ni

59.2

1.3

59.8

62.1

59.2

56.5

57.1

58.9

2.24

3.8%

100%

Pb

764

15

764

794

784

753

767

772

16.4

2.1%

101%

Sb

5.5

1.1

5.6

5.2

6

5.3

5.2

5.5

0.34

6.3%

99%

Se

0.78

0.14

0.82

0.86

0.78

0.8

0.71

0.79

0.055

7.0%

102%

V

27.2

1.4

26.6

26.9

26.1

26.0

26.0

26.3

0.409

1.6%

97%

Zn

1114

37

1130

1179

1156

1095

1100

1132

36

3.2%

102%

Click here to download the Application Note in PDF format.