SUMMARY
The application note summarizes the digestion of IARM 56H, an Inconel 718 Certified Reference Material by LGC ARMI , using ColdBlock Digestion Pro Series Technology.
Instrument: | ColdBlock CBM sample digester, chiller, HF compatible liners, ICP-OES |
Published: | February 2023 |
Digestion Time: | 20 Minutes |
Acid Used: | Aqua Regia, HF |
Average ColdBlock Recovery vs. CRM: |
|
Figure 1 – Inconel 718 after bulk-up to 50mL
IARM 56H was sourced from LGC ARMI in chips form. IARM-56H; Inconel 718 Alloy; UNS N07718; LGC ARMI,
Analytical Reference Materials International; Manchester, NH (January,2018)
Inconel 718 |
||||||||||
Method: |
0.25g |
20mL Aqua Regia + 1mL HF, Digested at 80% power for 20 minutes |
||||||||
Element |
LGC ARMI Certified Value wt.% |
Reference Value Limits |
Sample A |
Sample B |
Sample C |
Average (wt.%) |
Stdev |
% RSD |
% Recovery vs Certified Value |
|
Low |
High |
|||||||||
Al |
0.51 |
0.49 |
0.53 |
0.51 |
0.52 |
0.51 |
0.51 |
0.005 |
0.9% |
101% |
Cr |
17.86 |
17.8 |
17.92 |
17.74 |
17.91 |
17.86 |
17.84 |
0.071 |
0.4% |
100% |
Cu |
0.012 |
0.01 |
0.014 |
0.013 |
0.012 |
0.012 |
0.01 |
0.000 |
3.8% |
103% |
Fe |
18.6 |
18.5 |
18.7 |
18.9 |
18.4 |
18.7 |
18.67 |
0.205 |
1.1% |
100% |
Mn |
0.04 |
0.037 |
0.0043 |
0.04 |
0.04 |
0.04 |
0.04 |
0.000 |
1.2% |
97% |
Mo |
2.9 |
2.88 |
2.92 |
2.9 |
3.0 |
2.9 |
2.93 |
0.029 |
1.0% |
101% |
Nb |
5.2 |
5.17 |
5.23 |
5.2 |
5.1 |
5.0 |
5.12 |
0.095 |
1.9% |
98% |
Ni |
53.8 |
53.7 |
53.9 |
52.4 |
53.8 |
53.0 |
53.07 |
0.573 |
1.1% |
99% |
Si |
0.054 |
0.0024 |
0.057 |
0.059 |
0.057 |
0.051 |
0.06 |
0.003 |
6.1% |
103% |
Ta |
0.006 |
0.004 |
0.008 |
0.006 |
0.006 |
0.006 |
0.01 |
0.000 |
0.8% |
98% |
Ti |
0.96 |
0.95 |
0.97 |
0.95 |
0.96 |
0.96 |
0.96 |
0.005 |
0.5% |
100% |
Click here to download the Application Note in PDF format.